| Model | EC0A | EC01 | EC02 | EC03 |
|---|---|---|---|---|
| Working Volume | .37ft 3 (0.010 m3) |
.7ft 3 (0.020 m3) |
2.2 ft3 (0.062 m3) |
3.0 ft3 (0.085 m3) |
| Work Space Dim. [inches(cm)] |
10.0 (25.4) W 8.0 (20.3) H 8.0 (20.3) D |
12.0 (30.5) W 9.75 (24.75) H 10.25 (26.0) D |
20.0 (50.8) W 12.0 (30.5) H 16.0 (40.6) D |
20.0 (50.8) W 12.0 (30.5) H 22.0 (55.9) D |
| Overall Dim. [inches(cm)] |
17.5 (44.5) W 13.25 (33.7) H 20.0 (50.8) D |
21.0 (53.3) W 16.0 (40.6) H 24.0 (61.0) D |
30.0 (76.2) W 19.5 (49.5) H 30.0 (76.2) D |
30.0 (76.2) W 19.5 (49.5) H 36.0 (91.4) D |
| Max Rate |
.5ºC/sec (30ºC/min) |
.5ºC/sec (30ºC/min) |
.25ºC/sec (15ºC/min) |
.25ºC/sec (15ºC/min) |
| Air-Flow (CFM, Vertical) | 60 ft3/min | 60 ft3/min | 150 ft3/min | 300 ft3/min |
| Power 110VAC 50/60 Hz | YES | YES | YES | NO |
| Power 208/220VAC 50/60Hz Single Phase, 3/4 Wire |
YES | YES | YES | YES |
| Weight; Net/Shipping | 38/55 lbs | 49/68 lbs | 80/100 lbs | 98/115 lbs |
| Power Requirement | 1600 Watts | 1800 Watts | 2200 Watts | 3000 Watts |
|
Expendable Gas Coolant Usage LCO2 @ 0°C |
1.3 lbs/hr | 2 lbs/hr | 3.5 lbs/hr | 5 lbs/hr |
|
Expendable Gas Coolant Usage LCO2 @ -55°C |
4 lbs/hr | 6 lbs/hr | 11 lbs/hr | 15 lbs/hr |
|
Expendable Gas Coolant Usage LN2 @ 0°C |
.7 lbs/hr | 1 lbs/hr | 2 lbs/hr | 2.5 lbs/hr |
|
Expendable Gas Coolant Usage LN2 @ -55°C |
2.2 lbs/hr | 3 lbs/hr | 5.5 lbs/hr | 7.5 lbs/hr |
TC01 Specs (pdf)
TEMPERATURE
Set Temp. Range: -199ºC to +314ºC
Number of Scan Set Points: 10
Absolute Error Over Temp. Range: ± 1ºC
Temp. Control: ± .2ºC
Chamber Temp. Meast. Resolution: .1ºC
Temp. Display Res.: .1ºC less < 200ºC, 1ºC > 200ºC
Long Term Stability: ± .25ºC
Repeatablilty: ± .1ºC
Temp. Control Technique: PID
algorithm - PWM @ .5 Hz
TIME
Set Time-at Temperature Range 1-1800 minutes or continuous
Number of Scan Set Times: 10
Time-at-Temperature Resolution: .1 minute
CONTROL
Local Operation: 16 Key Keyboard / 4 Digit Display with red LED
time-out and remote indicator, audible tone.
Remote:
RS232-C 9600 Baud, IEEE-488 Bus
Interrupts: IEEE-488 SRQ INT
RS232 Interrupt ASCII "I"
Automatic Cycling of Programmed SCAN Time/Temp. Seq. 1-1800
cycles or continuous.
Additional I/O: 2 Auxiliary Outputs, 1 Auxiliary Input
Safety Features: Adjustable Over Temperature Thermostat,
Software Upper Temperature Limit.
Convenience Features: Heat and Cool Enable/Disable Switches