EC1x Summary Table (pdf)
Model EC1A EC11A EC10 EC11 EC12 EC13
Working Volume .37ft 3
(0.010 m3)
.37ft 3
(0.010 m3)
.7ft 3
(0.020 m3)
.7ft 3
(0.020 m3)
2.2 ft3
(0.062 m3)
3.0 ft3
(0.085 m3)
Work Space Dim. [inches(cm)] 10.0 (25.4) W
8.0 (20.3) H
8.0 (20.3) D
10.0 (25.4) W
8.0 (20.3) H
8.0 (20.3) D
12.0 (30.5) W
9.75 (24.75) H
10.25 (26.0) D
12.0 (30.5) W
9.75 (24.75) H
10.25 (26.0) D
20.0 (50.8) W
12.0 (30.5) H
16.0 (40.6) D
20.0 (50.8) W
12.0 (30.5) H
22.0 (55.9) D
Overall Dim. [inches(cm)] 17.5 (44.5) W
13.25 (33.7) H
20.0 (50.8) D
17.5 (44.5) W
13.25 (33.7) H
20.0 (50.8) D
21.0 (53.3) W
16.0 (40.6) H
24.0 (61.0) D
21.0 (53.3) W
16.0 (40.6) H
24.0 (61.0) D
30.0 (76.2) W
19.5 (49.5) H
30.0 (76.2) D
30.0 (76.2) W
19.5 (49.5) H
36.0 (91.4) D
Max Rate .5ºC/sec
(30ºC/min)
1.0ºC/sec
(60ºC/min)
.5ºC/sec
(30ºC/min)
.8ºC/sec
(48ºC/min)
.25ºC/sec
(15ºC/min)
.25ºC/sec
(15ºC/min)
Air-Flow (CFM, Vertical) 60 ft3/min 120 ft3/min 60 ft3/min 120 ft3/min 150 ft3/min 300 ft3/min
Power 110VAC 50/60 Hz YES NO YES NO YES NO
Power 208/220VAC
50/60Hz Single Phase, 3/4 Wire
YES YES YES YES YES YES
Weight; Net/Shipping 38/55 lbs 38/55 lbs 49/68 lbs 49/68 lbs 80/100 lbs 98/115 lbs
Power Requirement 1600 Watts 4200 Watts 1800 Watts 4200 Watts 2200 Watts 3000 Watts
Expendable Gas Coolant Usage
LCO2 @ 0°C
1.3 lbs/hr 1.3 lbs/hr 2 lbs/hr 2 lbs/hr 3.5 lbs/hr 5 lbs/hr
Expendable Gas Coolant Usage
LCO2 @ -55°C
4 lbs/hr 4 lbs/hr 6 lbs/hr 6 lbs/hr 11 lbs/hr 15 lbs/hr
Expendable Gas Coolant Usage
LN2 @ 0°C
.7 lbs/hr .7 lbs/hr 1 lbs/hr 1 lbs/hr 2 lbs/hr 2.5 lbs/hr
Expendable Gas Coolant Usage
LN2 @ -55°C
2.2 lbs/hr 2.2 lbs/hr 3 lbs/hr 3 lbs/hr 5.5 lbs/hr 7.5 lbs/hr
EC1x Summary Table (Continued)
Model EC13HA EC15HA EC16 EC16HA EC1.3W
Working Volume 3.0 ft3
(0.085 m3)
5.0 ft3
(0.142 m3)
6.16 ft3
(0.174 m3)
6.16 ft3
(0.174 m3)
1.37 ft3
(0.074 m3)
Work Space Dim. [inches(cm)] 20.0 (50.8) W
12.0 (30.5) H
22.0 (55.9) D
20.0 (50.8) W
12.0 (30.5) H
36.0 (91.4) D
22.0 (55.9) W
22.0 (55.9) H
22.0 (55.9) D
22.0 (55.9) W
22.0 (55.9) H
22.0 (55.9) D
14.0 (35.6) W
13.0 (33.0) H
13.0 (33.0) D
Overall Dim. [inches(cm)] 30.0 (76.2) W
19.5 (49.5) H
36.0 (91.4) D
30.0 (76.2) W
19.5 (49.5) H
50.0 (127.0) D
32.0 (81.3) W
29.5 (74.9) H
36.0 (91.4) D
32.0 (81.3) W
29.5 (74.9) H
36.0 (91.4) D
26.5 (67.3) W
21.0 (53.3) H
31.0 (78.7) D
Max Rate .5ºC/sec
(30ºC/min)
.5ºC/sec
(30ºC/min)
.4ºC/sec
(24ºC/min)
.4ºC/sec
(24ºC/min)
.33ºC/sec
(20ºC/min)
Air-Flow (CFM, Vertical) 600 ft3/min 600 ft3/min 600 ft3/min 600 ft3/min 250 ft3/min
Power 110VAC 50/60 Hz NO NO NO NO NO
Power 208/240VAC
50/60Hz Single Phase, 3/4 Wire
YES YES YES YES YES
Weight; Net/Shipping 108/125 lbs 138/175 lbs 139/160 lbs 165/185 lbs 77/94 lbs
Power Requirement 6000 Watts 6000 Watts 6000 Watts 6000 Watts 3000 Watts
Expendable Gas Coolant Usage
LCO2 @ 0°C
5 lbs/hr 10lbs/hr 12 lbs/hr 12 lbs/hr 3 lbs/hr
Expendable Gas Coolant Usage
LCO2 @ -55°C
15 lbs/hr 30 lbs/hr 35 lbs/hr 35 lbs/hr 10 lbs/hr
Expendable Gas Coolant Usage
LN2 @ 0°C
2.5 lbs/hr 5lbs/hr 6 lbs/hr 6 lbs/hr 2 lbs/hr
Expendable Gas Coolant Usage
LN2 @ -55°C
7.5 lbs/hr 15lbs/hr 18 lbs/hr 18 lbs/hr 5 lbs/hr

TC10/EC1x Controller Specs

Temperature Set Point Range: -200º C to +315ºC in 0.1ºC Steps
***EC1.3w -200ºC to +400ºC in 0.1ºC Steps***
Time at Temperature Range: 1.0 sec to 99 hr 59 min 59 sec or Continuous
Ramping Rate: 0.01ºC/min to chamber spec
Number of Set Temps/Times: Limited Only By Available Program Memory
(Typically 100+ Segments)
Number of Probe channels: 4 (2 for control, 2 additional read only upon request)
Absolute Error Over Temp Range (not including probe error): ±0.5ºC ( ±1ºC typical)
Long Term Stability (per month): ±0.2ºC
Temperature Resolution: 0.02ºC
Line Voltage Sensitivity (105-125 VAC): ±0.1ºC
Temperature Control Technique: PID Algorithm / Pulse Width Modulated
Local Operation: 29 Key Keyboard / 2 Line LCD
Remote Operation: Built-in IEEE-488 / RS232 / RS422 Interfaces
Analog output channels: 4 (0-5Vdc), 8 bit A/D
Analog input channels: 4 (0-5Vdc), 8 bit D/A, ±1 LSB accuracy
Auxiliary TTL output: Up to 5 available for external relay control
Auxiliary TTL input: 8 available for Logic High/Low sense
User Bus: 16 bit Address/ 8 bit data parallel port
Safety Features: Open/Short Probe Detection, Adjustable Over Temperature Thermostat or Electronic Limit Module, Software Upper and Lower Temperature Limit
Convenience Features: Configurable for Cooling options such as Auto Line Purge, 2 or 3 tank and redundant valves, Heat/Cool Enable/Disable Switches,Non-Volatile Memory for Programs, Setup Menus for Probe CAL, Remote Interrupt Operation and Default Conditions for Alarms and Analog and remote Interfaces, Time of Day Clock