Safety Features

Hardware

Over Temperature Protection

All of Sun's Temperature Cycling Chambers offer a failsafe protection mechanism to protect the chamber from thermal run-away due to hardware failure or operator error. Most of our chambers come standard with an easily adjustable (using a flat-head screwdriver), N.O (Normally Open), bi-metal thermostat that is wired to a breaker on the back panel of the chamber. When the air temperature inside the chamber exceeds the trip point of the thermostat, the breaker trips, removing all power from the heaters. The user is notified of this condition through a failsafe indicator/message on the chamber's display. Most of our chambers also have the option of having an Electronic Failsafe (Limit Module) installed in lieu of a bi-metal thermostat. The Limit Module, shown below, is mounted on the back of the chamber and continuously monitors the chamber air temperature. The limit module's temperature is continously displayed on the modules LCD. The module has both an adjustable upper and lower temperature limit. These limits are adjusted using a precision flat-head screwdriver and are displayed when the button for each is pressed. The RED button displays the upper temperature limit and the BLUE button displays the lower temperature limit. The module also has built-in GFI protection. The module will trip under three conditions: when there is an imbalance of current (>2A) to ground (GFI protection), when the limit module temperature exceeds the upper temperature limit setting and when the limit module temperature is less than the lower temperature limit setting.

over-temp-protection thermostat
Mechanical Failsafe (Standard)
Thermostat
over-temp-protection thermostat
Electronic Failsafe (Option)
Limit Module

Software

User adjustable upper and lower temperature limits

The TC10 controller features both an upper and lower user adjustable temperature limits that will prevent the operator of the chamber from inadvertently entering a set point higher or lower than these limits. These limits are acknowledged remotely and in local programs as well as through keyboard entry

A message will be displayed on the front panel "SET > UTL" for set points higher than the UTL and "SET < LTL" for set points that are lower than the LTL and the set point entered is ignored.

Door Sensor

An optional feature that can be installed on most EC1x models is a magnetic door sensor/switch. This sensor, when enabled in the SDEF menu, will check the status of the door before and during heating/cooling. If the door is opened during or before heating/cooling and a set point is entered, the controller will disable both heating and cooling and display a messsage on the LCD "B0 force H/C off". Once the door is closed, both heat and cool can be enabled by the user and heating/cooling will resume if a valid set point is entered.